日本菊水(KIKUSUI)TOS1200耐壓機電流校正儀詳細介紹:
菊水Kikusu TOS1200 耐壓機電流校正器Calibration of Leakage Current Detection Sensitivity Direct Reading of Error from Error Display Scale Ammeter Ranges Eliminates Need for Power Supply
菊水Kikusu TOS1200 耐壓機電流校正器 的詳細介紹
菊水Kikusu TOS1200 耐壓機電流校正器
Measuring Function | Measurement of current values and error (%) for AC (50/60 Hz) and DC at a test voltage of 1000 V |
Measuring Ranges | 8 ranges consisting of 0.5/1/2/5/10/20/50/100 mA along with values equal to 0.8 times the values of those ranges (for 1, 2, 4 and 8 steps) |
Ammeter Scale | Main scale: Direct-reading errordisplay scale over a range of ±10%of the above full scale values Auxiliary scale: Ratio scale of 0 to1.1 times the above full scale valuesAccuracy: (equivalent to 0% display of mainscale when the ratio is equal to 1)
|
Ammeter Accuracy | Main scale: ±1% of reading Auxiliary scale: ±3% of full scale value |
Ammeter Indication | DC/AC(sine wave rms value calibration of mean value response) |
Load Resistance | |||
Range[mA] | Resistance[k?] | Range[mA] | Resistance[k?] |
0.5 | 2000 | 10 | 100 |
1 | 1000 | 20 | 50 |
2 | 500 | 50 | 20 |
5 | 200 | 100 | 10 |
Allowed Input Time | 0.5/1/2/5 mA ranges: Continuous 10/20/50/100 mA ranges: 60 sec. Max. 1/3 of duty cycle | ||
Dimensions (MAX) | 134W × 164H × 270D mm (140W × 189H × 320D mm) | ||
Weight | approx. 3.5 kg | ||
Accessories | TL04-TOS high-voltage test lead: 1 |