欧美……一区二区三区,欧美日韩亚洲另类视频,亚洲国产欧美日韩中字,日本一区二区三区dvd视频在线

普升科技有限公司

多導(dǎo)生理記錄儀,微操作器,微電極拉制儀,生理信號(hào)記錄分析系統(tǒng),離體灌流系統(tǒng),腦立體定位儀,微量注射泵,眼動(dòng)儀,血流量計(jì),膜片鉗,細(xì)胞離子成像,迷宮和水迷宮,鼠尾無(wú)創(chuàng)血壓測(cè)量?jī)x,蠕動(dòng)泵,心功能測(cè)量,切片機(jī),動(dòng)物麻醉機(jī),電刺激器,凝膠圖像分析,藥效行為檢測(cè),血糖乳酸分析儀,動(dòng)物行為觀察分析系統(tǒng)

化工儀器網(wǎng)收藏該商鋪

21

聯(lián)系電話

18510328756

     小標(biāo) 您所在位置:首頁(yè) > 技術(shù)文章 > 700B Headstage Calibration
產(chǎn)品搜索

請(qǐng)輸入產(chǎn)品關(guān)鍵字:

多導(dǎo)生理記錄儀

微操作器

微電極拉制儀

細(xì)胞電生理

離體灌流系統(tǒng)

眼動(dòng)儀

膜片鉗

鼠尾無(wú)創(chuàng)血壓測(cè)量?jī)x

心功能測(cè)量

電刺激器

血糖乳酸分析儀

動(dòng)物行為觀察分析系統(tǒng)

微電極和電極陣列

情醒動(dòng)物生理信號(hào)遙測(cè)系統(tǒng)

DSI BUXCO 口鼻暴露吸入染毒系統(tǒng)

NAM小動(dòng)物肺功能-無(wú)創(chuàng)氣道機(jī)制檢測(cè)系統(tǒng)

小動(dòng)物肺功能-氣道阻力和肺順應(yīng)性系統(tǒng)

清醒實(shí)驗(yàn)動(dòng)物檢測(cè)系統(tǒng)

清醒小動(dòng)物呼吸肺功能檢測(cè)系統(tǒng)

心理人因產(chǎn)品

聯(lián)系方式
地址:北京市朝外大街乙12號(hào)昆泰國(guó)際大廈1806室
郵編:100020
聯(lián)系人:周經(jīng)理
電話:010-58790591
傳真:010-58790595
手機(jī):18510328756
留言:發(fā)送留言
網(wǎng)址:www.upwardstek.cn
商鋪:http://www.hbwxwy.cn/st3671/
技術(shù)文章

700B Headstage Calibration

點(diǎn)擊次數(shù):1075 發(fā)布時(shí)間:2015-1-28

700B Headstage Calibration

The steps below provide a quick check of the calibration of the MultiClamp 700B.  It is assumed that appropriate shielding (as described in “Test the Noise", above) will be used during these tests.

1.  Connect an oscilloscope to the front panel PRIMARY or SCOPE Output BNC.

2.  Synchronize the oscilloscope by connecting to the rear panel SYNC OUTPUT BNC. Press the “Reset to Program Defaults" button in the MultiClamp 700B Commander to standardize the MultiClamp 700B.


50 G Range

1.  Press the “Options" button, choose the Gains tab, and select the 50 G feedback resistor in the Voltage Clamp pane.  Return to the main MultiClamp 700B Commander window.

2.  Plug the PATCH connector of the PATCH-1U model cell into the CV-7 headstage.

3.  Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check the box to make it active.

4.  Press Auto Cp Fast to remove the bulk of electrode capacitance transient.

5.  The resulting waveform should be square, except for an initial overshoot (possibly twice the size of the steady-state response) that settles to the baseline in about 1 to 2 ms.  The rise time to the peak of the overshoot should be about 50 μs.  The steady-state amplitude following the transient should be ~500 mVp-p (±50 mV).


5 G Range

1.  Change the feedback resistor to 5 G in the Options / Gains menu.

2.  Press Auto Cp Fast.

3.  The step response should be ~50 mVp-p (±5 mV).


500 M Range

1.  Press the “Reset to Program Defaults" button.  (By default, the 500 M

range is selected.)

2.  Check Seal Test and set the amplitude to 25 mV.

3.  Plug the CELL connector of the PATCH-1U model cell into the CV-7 headstage.

4.  Press the Auto Whole Cell button.

5.  Press Auto Cp Fast button.

6.  The step response should be ~25 mVp-p. 


50 M Range

1.  Change the feedback resistor to 50 M.

2.  Increase Output Gain to 10.

3.  Press Auto Whole Cell and Auto Cp Fast.

4.  The step response should be ~25 mVp-p.  



[ 打印 ] [ 返回頂部 ] [ 關(guān)閉

| 商鋪首頁(yè) | 公司檔案 | 產(chǎn)品展示 | 供應(yīng)信息 | 公司動(dòng)態(tài) | 詢價(jià)留言 | 聯(lián)系我們 | 會(huì)員管理 |
化工儀器網(wǎng) 設(shè)計(jì)制作,未經(jīng)允許翻錄必究.Copyright(C) http://www.hbwxwy.cn, All rights reserved.
以上信息由企業(yè)自行提供,信息內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由相關(guān)企業(yè)負(fù)責(zé),化工儀器網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任。
溫馨提示:為規(guī)避購(gòu)買風(fēng)險(xiǎn),建議您在購(gòu)買產(chǎn)品前務(wù)必確認(rèn)供應(yīng)商資質(zhì)及產(chǎn)品質(zhì)量。
二維碼

掃一掃訪問手機(jī)站